172
6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
To use Sequence triggering
Sequence trigger lets you trigger the oscilloscope after finding a
sequence of events. Defining a sequence trigger requires three
steps:
1
Define an event to find before you search for the trigger.
The “find” event can be a pattern, an edge from a single
channel, or the combination of a pattern and a channel edge.
2
Define the trigger event.
The “trigger on” event can be a pattern, an edge from a single
channel, the combination of a pattern and a channel edge, or
the nth occurrence of an edge from a single channel.
3
Set an optional reset event.
If you choose to define a “reset” event, the event can be a
pattern, an edge from a single channel, the combination of a
pattern and a channel edge, or a timeout value.
Figure 14
Sequence Trigger Flowchart
To access the sequence trigger settings, press the
More
key in
the Trigger section of the front panel, rotate the Entry knob
until
Sequence
is displayed in the
Trigger
softkey, then press the
Settings
softkey to display Sequence trigger menu.
Trigger
holdoff
Scope
triggers
Start
Find:?
No
Yes
No
No
Yes
Yes
Trigger on:?
Reset on:?
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......