Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
197
As you press the
D+
or
D–
softkey (or rotate the Entry knob
on mixed-signal oscilloscopes), the
D+
and
D–
label for the
source channel is automatically set and the channel you
select is shown in the upper-right corner of the display next
to “
USB
”.
Adjust the trigger level for the selected analog channel by
turning the Trigger Level knob. Press the
D15 Thru D0
key and
select
Thresholds
to set the threshold level for digital
channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
4
Press the
Trigger:
softkey to select where the USB trigger will
occur:
•
SOP
(Start of Packet) — triggers at the Sync bit at the
beginning of the packet.
•
EOP
(End of Packet) — triggers at the end of the SE0
portion of the EOP.
•
RC
(Reset Complete) — triggers when SE0 is > 10 ms.
•
Suspend
(Enter Suspend) — triggers when the bus is idle
> 3 ms.
•
Exit Sus
(Exit Suspend) — triggers when exiting an idle
state > 10 ms. This is used to see the suspend/resume
transition.
D+,–
D–,+
Bus Idle
SOP
SE0 portion of EOP
End of packet trigger
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......