![Agilent Technologies InfiniiVision DSO6014L Скачать руководство пользователя страница 104](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-dso6014l/infiniivision-dso6014l_user-manual_2868144104.webp)
Front-Panel Controls
2
6000 Series Oscilloscope User’s Guide
103
Performing User Cal will invalidate your Certificate of
Calibration. If NIST (National Institute of Standards and
Technology) traceability is required, perform the Performance
Verification procedure in the
Agilent 6000 Series Oscilloscopes
Service Guide
using traceable sources.
To perform User Cal
1
Set the rear-panel CALIBRATION switch to UNPROTECTED.
2
Connect short (12 inch maximum) equal length cables to
each analog channel’s BNC connector on the front of the
oscilloscope. You will need two equal-length cables for a
2-channel oscilloscope or four equal-length cables for a
4-channel oscilloscope.
Use 50
Ω
RG58AU or equivalent BNC cables when performing
User Cal.
For a 2-channel oscilloscope, connect a BNC tee to the equal
length cables. Then connect a BNC(f)-to-BNC(f) (also called a
barrel connector) to the tee as shown below.
Figure 10
User Calibration cable for 2-channel oscilloscope
Longer cable
to TRIG OUT
To Channel 1
To Channel 2
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......