780
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
36
Synchronizing Acquisitions
Synchronization in the Programming Flow
Most remote programming follows these three general steps:
1
Set up the oscilloscope and device under test (see
page 780
).
2
Acquire a waveform (see
page 780
).
3
Retrieve results (see
page 780
).
Set Up the Oscilloscope
Before making changes to the oscilloscope setup, it is best to make sure it
is stopped using the :STOP command followed by the *OPC? query.
Acquire a Waveform
When acquiring a waveform there are two possible methods used to wait
for the acquisition to complete. These methods are blocking and polling.
The table below details when each method should be chosen and why.
Retrieve Results
Once the acquisition is complete, it is safe to retrieve measurements and
statistics.
N O T E
It is not necessary to use *OPC?, hard coded waits, or status checking when setting up the
oscilloscope. After the oscilloscope is configured, it is ready for an acquisition.
Blocking Wait
Polling Wait
Use When
You know the oscilloscope will
trigger based on the oscilloscope
setup and device under test.
You know the oscilloscope may or
may not trigger on the oscilloscope
setup and device under test.
Advantages
No need for polling.
Fastest method.
Remote interface will not timeout
No need for device clear if no trigger.
Disadvantages
Remote interface may timeout.
Device clear only way to get control
of oscilloscope if there is no trigger.
Slower method.
Requires polling loop.
Requires known maximum wait time.
Implementation
Details
See
"Blocking Synchronization"
on
page 781.
See
"Polling Synchronization With
Timeout"
on page 782.
Содержание InfiniiVision 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 120: ...120 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Страница 276: ...276 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 15 EXTernal Trigger Commands...
Страница 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Страница 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Страница 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Страница 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Страница 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Страница 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Страница 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Страница 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...
Страница 788: ...788 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 36 Synchronizing Acquisitions...
Страница 918: ...918 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 38 Programming Examples...