:TRIGger Commands
29
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
599
:TRIGger:NREJect
(see
page 790
)
Command Syntax
:TRIGger:NREJect <value>
<value> ::= {{0 | OFF} | {1 | ON}}
The :TRIGger:NREJect command turns the noise reject filter off and on.
When the noise reject filter is on, the trigger circuitry is less sensitive to
noise but may require a greater amplitude waveform to trigger the
oscilloscope. This command is not valid in TV trigger mode.
Query Syntax
:TRIGger:NREJect?
The :TRIGger:NREJect? query returns the current noise reject filter mode.
Return Format
<value><NL>
<value> ::= {0 | 1}
See Also
•
"Introduction to :TRIGger Commands"
on page 589
Содержание InfiniiVision 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 120: ...120 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Страница 276: ...276 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 15 EXTernal Trigger Commands...
Страница 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Страница 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Страница 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Страница 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Страница 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Страница 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Страница 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Страница 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...
Страница 788: ...788 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 36 Synchronizing Acquisitions...
Страница 918: ...918 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 38 Programming Examples...