264
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
14
:DVM Commands
:DVM:ARANge
(see
page 790
)
Command Syntax
:DVM:ARANge <setting>
<setting> ::= {{OFF | 0} | {ON | 1}}
If the selected digital voltmeter (DVM) source channel is not used in
oscilloscope triggering, the :DVM:ARANge command turns the digital
voltmeter's Auto Range capability on or off.
•
When on, the DVM channel's vertical scale, vertical (ground level)
position, and trigger (threshold voltage) level (used for the counter
frequency measurement) are automatically adjusted.
The Auto Range capability overrides attempted adjustments of the
channel's vertical scale and position.
•
When off, you can adjust the channel's vertical scale and position
normally.
Query Syntax
:DVM:ARANge?
The :DVM:ARANge? query returns a flag indicating whether the digital
voltmeter's Auto Range capability is on or off.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also
•
":DVM:SOURce"
on page 269
•
":DVM:ENABle"
on page 266
•
":DVM:MODE"
on page 268
Содержание InfiniiVision 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 120: ...120 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Страница 276: ...276 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 15 EXTernal Trigger Commands...
Страница 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Страница 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Страница 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Страница 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Страница 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Страница 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Страница 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Страница 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...
Страница 788: ...788 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 36 Synchronizing Acquisitions...
Страница 918: ...918 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 38 Programming Examples...