:SBUS<n> Commands
25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
529
:SBUS<n>:UART:TRIGger:TYPE
(see
page 790
)
Command Syntax
:SBUS<n>:UART:TRIGger:TYPE <value>
<value> ::= {RSTArt | RSTOp | RDATa | RD1 | RD0 | RDX | PARityerror
| TSTArt | TSTOp | TDATa | TD1 | TD0 | TDX}
The :SBUS<n>:UART:TRIGger:TYPE command selects the UART trigger type.
When one of the RD or TD types is selected, the
:SBUS<n>:UART:TRIGger:DATA and :SBUS<n>:UART:TRIGger:QUALifier
commands are used to specify the data value and comparison operator.
The RD1, RD0, RDX, TD1, TD0, and TDX types (for triggering on data and
alert bit values) are only valid when a 9- bit width has been selected.
Query Syntax
:SBUS<n>:UART:TRIGger:TYPE?
The :SBUS<n>:UART:TRIGger:TYPE? query returns the current UART
trigger data value.
Return Format
<value><NL>
<value> ::= {RSTA | RSTO | RDAT | RD1 | RD0 | RDX | PAR | TSTA |
TSTO | TDAT | TD1 | TD0 | TDX}
See Also
•
"Introduction to :TRIGger Commands"
on page 589
•
":TRIGger:MODE"
on page 598
•
":SBUS<n>:UART:TRIGger:DATA"
on page 526
•
":SBUS<n>:UART:TRIGger:QUALifier"
on page 528
•
":SBUS<n>:UART:WIDTh"
on page 530
Содержание InfiniiVision 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 120: ...120 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Страница 276: ...276 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 15 EXTernal Trigger Commands...
Страница 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Страница 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Страница 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Страница 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Страница 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Страница 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Страница 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Страница 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...
Страница 788: ...788 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 36 Synchronizing Acquisitions...
Страница 918: ...918 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 38 Programming Examples...