674
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
31
:WGEN Commands
:WGEN:MODulation:FM:DEViation
(see
page 790
)
Command Syntax
:WGEN:MODulation:FM:DEViation <frequency>
<frequency> ::= frequency deviation in Hz in NR3 format
The :WGEN:MODulation:FM:DEViation command specifies the frequency
deviation from the original carrier signal frequency.
When the modulating signal is at its maximum amplitude, the output
frequency is the carrier signal frequency plus the deviation amount, and
when the modulating signal is at its minimum amplitude, the output
frequency is the carrier signal frequency minus the deviation amount.
The frequency deviation cannot be greater than the original carrier signal
frequency.
Also, the sum of the original carrier signal frequency and the frequency
deviation must be less than or equal to the maximum frequency for the
selected waveform generator function plus 100 kHz.
Query Syntax
:WGEN:MODulation:FM:DEViation?
The :WGEN:MODulation:FM:DEViation? query returns the frequency
deviation setting.
Return Format
<frequency><NL>
<frequency> ::= frequency deviation in Hz in NR3 format
See Also
•
":WGEN:MODulation:AM:DEPTh"
on page 672
•
":WGEN:MODulation:AM:FREQuency"
on page 673
•
":WGEN:MODulation:FM:FREQuency"
on page 675
•
":WGEN:MODulation:FSKey:FREQuency"
on page 676
•
":WGEN:MODulation:FSKey:RATE"
on page 677
•
":WGEN:MODulation:FUNCtion"
on page 678
•
":WGEN:MODulation:FUNCtion:RAMP:SYMMetry"
on page 679
•
":WGEN:MODulation:STATe"
on page 681
•
":WGEN:MODulation:TYPE"
on page 682
Содержание InfiniiVision 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 120: ...120 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Страница 276: ...276 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 15 EXTernal Trigger Commands...
Страница 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Страница 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Страница 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Страница 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Страница 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Страница 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Страница 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Страница 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...
Страница 788: ...788 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 36 Synchronizing Acquisitions...
Страница 918: ...918 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 38 Programming Examples...