:MEASure Commands
:MEASure:NWIDth
(see
page 790
)
Command Syntax
:MEASure:NWIDth [<source>]
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion | MATH
| WMEMory<r>}
<digital channels> ::= DIGital<d> for the MSO models
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :MEASure:NWIDth command installs a screen measurement and starts
a negative pulse width measurement. If the optional source parameter is
not specified, the current source is modified.
Query Syntax
:MEASure:NWIDth? [<source>]
The :MEASure:NWIDth? query measures and outputs the width of the
negative pulse on the screen closest to the trigger reference using the
midpoint between the upper and lower thresholds.
FOR the negative pulse closest to the trigger point:
width = (time at trailing rising edge - time at leading falling edge)
Return Format
<value><NL>
<value> ::= negative pulse width in seconds in NR3 format
See Also
•
"Introduction to :MEASure Commands"
on page 340
•
":MEASure:SOURce"
on page 361
•
":MEASure:PWIDth"
on page 358
•
":MEASure:PERiod"
on page 355
N O T E
This command is not available if the source is FFT (Fast Fourier Transform).
Содержание InfiniiVision 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Страница 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 120: ...120 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Страница 276: ...276 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 15 EXTernal Trigger Commands...
Страница 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Страница 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Страница 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Страница 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Страница 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Страница 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Страница 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Страница 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...
Страница 788: ...788 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 36 Synchronizing Acquisitions...
Страница 918: ...918 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 38 Programming Examples...