Installing the Autosampler
Installing the Thermostatted Autosampler
13
If an Agilent Chemstation is the controller, connect either
❏
The GPIB cable to the detector
❏
The LAN connector to the LAN interface
14
Connect the APG remote cable (optional) for non Agilent 1100 series
instruments.
15
Turn ON power by pushing the button at the lower left hand side of the
sampler.
N O T E
The sampler is turned ON when the line power switch is pressed and the
green indicator lamp is illuminated. The detector is turned OFF when the
line power switch is protruding and the green light is OFF.
WA R N I N G
To disconnect the sampler from the line, unplug the power cord. The
power supply still uses some power, even if the power switch at the front
of the panel is turned OFF.
WA R N I N G
To avoid personal injury, keep fingers away from the needle area during
Autosampler operation. Do not attempt to insert or remove a vial or a
plate when the needle is positioned.
Figure 9
Cable Connections
GPIB
RS232
CAN-bu
Remote
Vial number
Relay
Thermostat -
Autosampler cable
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......