14
Installing the Autosampler
Site Requirements
A suitable site environment is important to ensure optimal performance of
the autosampler.
Power Consideration
The autosampler power supply has wide-ranging capability. Consequently
there is no voltage selector in the rear of the autosampler. There are also
no externally accessible fuses, because automatic electronic fuses are
implemented in the power supply.
The thermostatted autosampler comprises two modules, the autosampler
and the thermostat (G1330B). Both modules have a separate power supply
and a power plug for the line connections. The two modules are connected
by a control cable and both are turned on by the autosampler module.
WA R N I N G
To disconnect the autosampler from line power, unplug the power cord.
The power supply still uses some power, even if the power switch on the
front panel is turned off.
WA R N I N G
To disconnect the thermostatted autosampler from line power, unplug
the power cord from the autosampler and the ALS thermostat. The
power supplies still use some power, even if the power switch on the
front panel is turned off. Please make sure that it is always possible to
access the power plug.
WA R N I N G
Shock hazard or damage of your instrumentation can result if the
devices are connected to a line voltage higher than specified.
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......