95
Troubleshooting and Test Functions
Intermittent lock-ups with or without vial in the gripper fingers
Intermittent lock-ups with or without vial
in the gripper fingers
With error messages
❏
motor overtemp (0 or 1 or 2 or 3)
❏
movement failed (0 or 1 or 2 or 3)
❏
missing vial
N O T E
When a motor over temperature message has occurred, the sampler must be
turned off for about 10 minutes to allow the motor to cool down.
Step 1
Check the firmware and update to the latest revision if necessary.
Since firmware revision A.03.61 (resident A03.60) most “movement failed”,
“motor over temp”, “initialization failed (X-axis)” errors are solved.
Step 2
Check the vials and the caps.
For reliable operation, vials used with the 1100 Autosampler must not have
tapered shoulders or caps that are wider than the body of the vial. For more
details see the service note G1313-017.
Step 3
Very heavy usage - use a macro.
A pre-sequence macro,
QMBUVHW_PDF
will automatically reset the
sampler at the start of a sequence (ChemStation).
Step 4
Check if the “INJECT” line is used in the “Injector Program”.
Remove this line from the program. In this mode the system does not need
this command to do the injection. A firmware revision (>3.81) will address
this problem. For more details see the service note G1313-018.
Step5
Reset the sampler alignment to default value.
Reset tray alignment, and transport alignment is possible with the Control
Module and the ChemStation. To reset the transport alignment with the
ChemStation, enter following command in the command line. Print
sendmodule$(lals, “tray:alig 0.00,0.00”)
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......