87
Troubleshooting and Test Functions
Tray Alignment
Tray Alignment
Tray alignment is required to compensate for small deviations in positioning
of the gripper which may occur after disassembling the module for repair.
The tray alignment procedure uses several tray positions as reference points.
Because the tray is a rectangle, a two-point alignment is sufficient to corrects
all other vial positions within the tray. On completion of the procedure, the
corrected gripper positions are stored in the instrument firmware.
N O T E
The alignment procedure requires an 1100 Control Module G1323B with
firmware revision B.02.02 or higher. The screen with the alignment dialog box
can be found under the menus
Views/System/Tests/Autosampler.
The
alignment procedure must be done with the standard 100-position vial tray
installed.
WA R NI N G
The alignment procedure has to be performed in the correct order and
without skipping parts, to ensure a proper working Autosampler.
User Interface
In the Control Module the “Align Tray” function is located in the Control
display of the autosampler.
Additional information can be found in the on-line information systems.
The correct procedure for the Control Module are:
1
Set alignment to factory default:
•
Go to
Align/Tray
and press button
Default
.
•
Go to
Align/Transport
and press button
Default
.
•
Wait while the Autosampler performs a reset and go back to
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......