232
Theory of Operation
Interfaces
Analog Signal Output
The analog signal output can be distributed to a recording device. For
details refer to the description of the module’s main board.
GPIB Interface
The GPIB connector is used to connect the module with a computer. The
address and control switches next to the GPIB connector determine the
GPIB address of your module. The switches are preset to a default address
and recognized by the operating software from Agilent Technologies.
CAN Interface
The CAN is an intermodule communication interface. It is a 2-wire serial
bus system supporting high speed data communication and real-time
requirement.
Remote Interface
The APG remote connector may be used in combination with other
analytical instruments from Agilent Technologies if you want to use
features such as common shut down, prepare, and so on.
Remote control allows easy connection between single instruments or
systems to ensure coordinated analysis with simple coupling
requirements.
The subminiature D connector is used. The module provides one remote
connector which is inputs/outputs (wired-or technique).
Table 50
Default Addresses
Autosampler
28
Autosampler
28
Pump
22
RID
29
FLD
23
VWD
24
Autosampler (HP 1050) 18
Agilent 8453A
25
Pump (HP 1050)
16
DAD
/ MWD
26
VWD (HP 1050)
10
Column Compartment
27
DAD (HP 1050)
17
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......