272
Specifications
Performance Specifications
Table 64
Performance Specifications Agilent 1100 Series Thermostatted Micro Autosampler
(G1389A)
7\SH
6SHFLILFDWLRQ
Sample capacity
100 x 2 ml vials in 1 tray. Microvials (100 or 300 µl) with
sleeves (reduced cooling performance with microvials)
Settable injection
volume
0.01 to 8 µl with small loop capillary
0.01 to 40 µl with extended loop capillary
Precision
Typically < 0.5 % RSD from 5 to 40 µl
Typically < 1 % RSD from 1 to 5 µl
Typically < 3 % RSD from 0.2 to 1 µl
Minimum sample
volume
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from
10 µl sample in 300 µl microvials
Carryover
Typically < 0.1 % without automated needle wash
Typically < 0.05 % with external needle cleaning and 1 µl
injection volume
Sample viscosity range
0.2 - 5 cp
Recommended
pH-range
1.0 - 8.5 solvent with pH < 2.3 should not contain acids
which attack stainless steel. Upper pH range is limited by
fused silica capillaries
Material in contact with
solvent
Stainless steel, sapphire, PTFE, PEEK, fused silica,
Vespel
GLP features
Early maintenance feedback (EMF), electronic records of
maintenance and errors
Communications
Controller-area network (CAN). GPIB (IEEE-448),
RS232C, APG-remote standard, optional four external
contact closures and BCD vial number output
Safety features
Leak detection and safe leak handling, low voltages in
maintenance areas, error detection and display
Housing
All material recyclable
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......