28
Installing the Autosampler
Installing the Thermostatted Autosampler
Installing the Thermostatted Autosampler
1
Place the Thermostat on the bench.
2
Remove the front cover and route the condensation drain tube to the waste.
WA R N I N G
Make sure that the condensation tube is always above the liquid level in
the vessel. If the tube is located in liquid the condensed water cannot
flow out of the tube and the outlet is blocked. Any further condensation
will then remain in the instrument. This may damage the instruments
electronics.
Figure 7
Condensation Leak outlet
3
Install the LAN interface board in the sampler (if required), see “Agilent
1100 Series Interface Board to general purposes” on page 198.
4
Remove the adhesive tape which covers the front door.
5
Remove the front door and remove the transport protection foam.
Preparation
Locate bench space
Provide power connection
Unpack the Sampler and the Thermostat
Parts required
Sampler and Thermostat
Power cord, for the other cables see below and “Cable Overview” on page 185
Chemstation and/or Control Module G1323B.
Condensation drain tube
Waste bottle
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......