Repairing the Autosampler
ASM Board
7
Install the transport assembly (page 132).
8
Install the top cover and foam (page 128).
9
Turn on the autosampler.
10
Enter the 10-character autosampler serial number. The serial number can be
entered using either the control module or the ChemStation, see “Entering the
Serial Number using the Control Module” on page 150 or see“Entering the
Serial Number using the ChemStation” on page 151.
11
Check the firmware revision of the autosampler. If the firmware revision is
older than the current firmware revision of the autosampler, update the
firmware, see “Replacing the Autosampler Firmware” on page 151
Entering the Serial Number using the Control Module
1
Connect the control module to the autosampler. Turn on the autosampler.
2
In the control module, press
System (F5)
, then
Records (F4)
. Using the
up/down arrows, make sure that the autosampler is highlighted.
3
Press
FW Update (F5)
. Now, press the
m
key. This will display a box which
says ‘
Update Enter Serial#’
.
4
Press
Enter
. This will display the box labeled
Serial#
.
5
Letters and numbers are created using the up and down arrows. Into the box
labeled
Serial#
, enter the 10-character serial number for the autosampler.
When the 10-character serial number is entered, press
Enter
to highlight the
complete serial number. Then, press
Done (F6)
.
N O T E
For firmware revisions below A02.00 it is very important never to press
Done
if the Serial# box is blank. In this case, the module can no longer be recognized
by either the control module or the ChemStation. The main board must then
be replaced.
6
Turn the autosampler off, then on again. The
Records
screen should display
the correct serial number for this module.
7
If a ChemStation is also connected, re-boot the ChemStation now as well.
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......