Service Guide E8364-90026
3-49
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362B, E8363B, E8364B
Adjustments
Receiver Calibration Adjustment
The receiver calibration is used to adjust the network analyzer receivers for a flat response
across its full frequency range:
1. A power meter/sensor is connected to Port 1, as shown in
, to establish a
reference for flatness.
2. A cable is inserted between the power sensor and the test port, as shown in
,
to establish a reference for the cable.
3. The same cable is connected between the two test ports, as shown in
, and a
signal from Port 1 is used to adjust the “B” receiver at Port 2.
The adjustment is repeated using a signal from Port 2 to adjust the “A” receiver at
Port 1.
Data obtained during this adjustment are stored in the
mxcalfile_pxx
files on the hard disk
drive (where
xx
is a 1 or 2 digit number). The data are used in subsequent measurements.
If the hard disk drive is replaced, these
mxcalfile_pxx
files will be lost. Therefore, they
should be backed up (saved on a floppy disk) so that they can be restored. If using multiple
disk drives (e.g. classified and general usage), then these files must be replaced on each
individual disk drive.
These files can be recreated by performing another receiver calibration adjustment.
Equipment Used for the Receiver Calibration Adjustment
Procedure
1. Connect the equipment as shown in
. Connect a GPIB cable between the
network analyzer and the power meter.
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
Receiver Calibration
.
3. Ensure the GPIB settings are correct.
4. Click
Calibrate
, and then follow the instructions as they are displayed.
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model
Part Number
Power meter
All
E4418B/E4419B
E4418A/E4419A
Power sensor, 3.5 mm
E8362B
E4413A
8485A
Adapter, 3.5 mm (f) to 3.5 mm (f)
E8362B
83059B
85052-60012
RF Cable, 3.5 mm (f) to 3.5 mm (f)
E8362B
85131C
85131E
Power sensor, 2.4 mm
E8363B, E8364B
8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
E8363B, E8364B
11900B
85056-60007
RF Cable, 2.4 mm (f) to 2.4 mm (f)
E8363B, E8364B
85133C
85133E
Содержание E8362B
Страница 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Страница 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Страница 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Страница 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Страница 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Страница 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Страница 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...