3-24
Service Guide E8364-90026
Tests and Adjustments
PNA Series Microwave Network Analyzers
System Verification
E8362B, E8363B, E8364B
Precautions for Handling Airlines
The verification kit includes airlines. Follow the steps below to ensure good measurement
techniques.
• Be very careful not to drop the airline’s center or outer conductor. Irreparable damage
will result if these devices are dropped.
• During this procedure, you will be touching the exposed center conductor of the test
port and the center conductor of the airline.
— Observe proper ESD procedures.
— Clean your hands or wear gloves; skin oils will cause a change in electrical
performance.
,
, and to your calibration kit manual for connection
procedures.
Figure 3-9
Aligning the Center Conductor
Figure 3-10 Torquing the Connection