Service Guide E8364-90026
3-11
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362B, E8363B, E8364B
Preliminary Checks
Preliminary Checks
Preliminary checks include the following:
•
“The Operator’s Check” on page 3-11
The operator’s check tests the network analyzer’s basic functionality of the source,
switch, step attenuators, and receivers.
•
“The Test Port Cable Checks” on page 3-13
The test port cable checks are not required, but are recommended to verify the
performance of the test port cables before performing the verification test.
The Operator’s Check
NOTE
To achieve the maximum system stability, allow the analyzer to warm up for
at least 90 minutes before performing the Operator’s Check.
The operator’s check is a software driven test that checks the basic operation of the
assemblies in the Port 1 and Port 2 signal paths. By performing the operator’s check, the
following are determined:
• repeatability of the RF test port switch
• attenuation ranges of all installed attenuators
• calibration of the receivers
• frequency response of the receivers
• phase lock and leveling
• noise floor and trace noise
Accessories Used in the Operator’s Check
Performing the Operator’s Check
1. From the
System
menu, point to
Service
, and then click
Operator’s Check
.
2. In the
PNA
Operator’s Check
dialog box (refer to
), under
Configure
, select
either
Prompt for attachment of Short/Open
, to pause at each step in the process to
allow moving the short/open to the appropriate port, or
Shorts/Opens are attached to
ALL ports
, to run through the test without stopping. Shorts and opens can be mixed on
the test ports.
Equipment Type
Models Used With
Part Number
Female short, 3.5 mm
E8362B
85052-60007 (from the 85052 calibration kit)
Female open, 3.5 mm
E8362B
85052-60009 (from the 85052 calibration kit)
Female short, 2.4 mm
E8363B, E8364B
85056-60021 (from the 85056 calibration kit)
Female open 2.4 mm
E8363B, E8364B
85056-60023 (from the 85056 calibration kit)
Содержание E8362B
Страница 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Страница 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Страница 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Страница 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Страница 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Страница 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Страница 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...