
Service Guide E8364-90026
3-41
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362B, E8363B, E8364B
Adjustments
10 MHz Frequency Reference Adjustment
The 10 MHz frequency adjustment is used to adjust the frequency accuracy of the network
analyzer’s 10 MHz frequency reference on the A10 frequency reference board assembly.
Equipment Used for the Frequency Adjustment at 10 MHz
Procedure
NOTE
The specification of frequency accuracy is
±
1 ppm. However, this test typically
adjusts to
±
0.01 ppm.
1. Connect the equipment as shown in
. Connect a GPIB cable between the
network analyzer and the frequency counter.
Figure 3-17 Setup for Adjustment of the 10 MHz Frequency Reference
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
10 MHz Freq Adjust
.
3. Ensure the GPIB settings are correct.
4. Click
Begin Adj
, and then follow the instructions as they are displayed.
Equipment Type
Models
Used With
Model or
Part Number
Alternate Model
or Part Number
Cable, BNC, 50
Ω
, 24 inch
All
8120-1839
Any
Frequency counter
All
53151A
Opt 001
Any
a
a. Must be capable of measuring a signal at 10 MHz.
Содержание E8362B
Страница 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Страница 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Страница 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Страница 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Страница 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Страница 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Страница 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...