3-38
Service Guide E8364-90026
Tests and Adjustments
PNA Series Microwave Network Analyzers
Performance Tests (Agilent N7840A Software Package)
E8362B, E8363B, E8364B
Dynamic Accuracy Test
Function of the Test:
To measure the relative power linearity of the analyzer’s receivers.
Specification Tested:
Test Port Input–Dynamic Accuracy
Equipment Used:
The dynamic accuracy test set (Z5623A Option H01), a power meter
(E4418B or E4419B) power sensors, and two test cables.
Description of the Test:
1. The analyzer’s test ports are tested separately at a specific CW frequency and a
reference power level of
−
20 dBm.
2. The analyzer’s driving port is connected to the dynamic accuracy test set’s source port
and the analyzer’s receiving port is connected to the dynamic accuracy test set’s
receiver port. A power sensor is connected to the dynamic accuracy test set’s power
meter port.
3. The dynamic accuracy test set is used to input a signal from the driving port on the
analyzer. The input signal is routed through step attenuators to both the analyzer and
a power sensor.
4. The analyzer source and the dynamic accuracy test set create power levels that are
“deltas” from the reference power level of
−
20 dBm. Each delta is measured two ways:
by the power sensor and by the analyzer’s receiver under test.
5. The analyzer’s power level into the port under test is set to the
−
20 dBm reference
level.
6. The power level is measured with the power sensor: (
P
mr
).
7. The power level is measured using the analyzer’s receiver under test: (
P
pr
).
8. The power level into the analyzer’s port under test is stepped, in 5 dB steps, over the
range of 0 to
−
120 dBm. Each power level represents a “delta” from the reference level.
9. At each step, the power level is measured using the power meter: (
P
md
).
10. At each step, the power level is measured using the receiver under test: (
P
pd
).
11. The power error in dB
=
(
P
mr
−
P
md
)
−
(P
pr
−
P
pd
)
.
Содержание E8362B
Страница 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Страница 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Страница 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Страница 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Страница 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Страница 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Страница 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...