Service Guide E8364-90026
7-3
PNA Series Microwave Network Analyzers
Repair and Replacement Procedures
E8362B, E8363B, E8364B
Before Starting the Removal and Replacement Procedures
Before Starting the Removal and Replacement Procedures
Before starting the removal and replacement procedures, review these warnings, cautions,
and the assembly replacement sequence.
Personal Safety Warnings
WARNING
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you
are qualified to do so.
WARNING
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the analyzer from all voltage sources
while it is being opened.
WARNING
The power cord is connected to internal capacitors that may remain
live for 10 seconds after disconnecting the plug from its power
supply assembly. Wait at least 10 seconds, after disconnecting the
plug, before removing the covers.
Electrostatic Discharge (ESD) Protection
CAUTION
Many of the assemblies in this instrument are very susceptible to damage
from electrostatic discharge (ESD). Perform the following procedures only at
a static-safe workstation and wear a grounded wrist strap.
This is important. If not properly protected against, electrostatic discharge
can seriously damage your analyzer, resulting in costly repair.
To reduce the chance of electrostatic discharge, follow all of the
recommendations outlined in
“Electrostatic Discharge Protection” on page
, for all of the procedures in this chapter.
Assembly Replacement Sequence
The following steps show the sequence that you should follow to replace an assembly in the
network analyzer.
Step 1.
Identify the faulty group. Begin with
Step 2.
Order a replacement assembly. Refer to
Chapter 6, “Replaceable Parts.”
Step 3.
Replace the faulty assembly and determine what adjustments are necessary.
Refer to
in this chapter.
Step 4.
Perform the necessary adjustments. Refer to
Chapter 3, “Tests and Adjustments.”
Step 5.
Perform the necessary performance tests. Refer to
Содержание E8362B
Страница 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Страница 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Страница 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Страница 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Страница 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Страница 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Страница 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...