Chapter 3
73
Basic Operation
Using the Instrument State Registers
6. Press
Load Segment From NVWFM Memory
>
Return
.
7. Scroll to the waveform selected in step 3.
8. Press
Select Waveform
>
ARB Off On
to on. This causes the signal generator to play the waveform.
Saving the Instrument State
Instrument states can be saved to any one of the 10 sequences and 100 registers in instrument state memory.
In this procedure, sequence 01 and register 02 are used to store the signal generator settings associated with
the waveform file. The Save function saves instrument settings and the waveform file name but does not
save waveform data. Waveform data can only be saved to NVWFM with the Store function.
1. Press
Save
>
Select Seq
>
01
>
Select Reg
>
02
>
Save Reg
.
The signal generator settings associated with the waveform file and the name of the waveform file are now
saved in the instrument state memory. The section
“Recalling an Instrument State for a Waveform File” on
describes how to recall this instrument state and associated waveform file.
Recalling an Instrument State for a Waveform File
This procedure applies to signal generators with Option 001/601 or 002/602 and demonstrates recalling
signal generator settings associated with a waveform file. This procedure uses the same signal generator
settings and waveform file used in the previous section,
“Saving an Instrument State for a Waveform File”
1. Press
Mode > Dual ARB > Waveform Segments
.
2. Scroll and select the waveform file used in the previous section.
3. Press
Load Segment From NVWFM Memory
>
Return
>
Select Waveform
4. Highlight the file selected in step 3.
5. Press
Select Waveform
.
The next step recalls the signal generator settings associated with the waveform file. These settings were
saved to Seq 01 and Reg 02 in the previous procedure.
6. Press
Recall > Select Seq
>
01
7. Press
RECALL Reg > 01 > Enter
.
The signal generator is now setup with the same settings associated with the waveform previously saved to
NVWFM.
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...