Chapter 3
79
Basic Operation
Using Security Functions
Erase All
CAUTION
Do not use this function with signal generator firmware revisions C.04.84, C.04.86, or
C.04.95. If you have one of these revisions installed, please upgrade
immediately
to revision
C.04.96 or later.
This function removes all user files, user flatness calibrations, user I/Q calibrations, and resets all table
editors with original factory values, ensuring that user data and configurations are not accessible or
viewable. The instrument appears as if it is in its original factory state, however, the memory is not sanitized.
This action is relatively quick, taking less than one minute (the exact time depends on the number of files).
To carry out this function, press
Utility
>
Memory Catalog
>
More (1 of 2)
>
Security
>
Erase All
>
Confirm Erase
.
NOTE
This function is different than pressing
Utility
>
Memory Catalog
>
More (1 of 2)
>
Delete All Files
, which deletes all user files, but does not reset the table editors.
Erase and Overwrite All
CAUTION
Do not use this function with signal generator firmware revisions C.04.84, C.04.86, or
C.04.95. If you have one of these revisions installed, please upgrade
immediately
to revision
C.04.96 or later.
This function performs the same actions as
and then clears and overwrites the various memory
types in accordance with Department of Defense (DoD) standards, as described below.
To carry out this function, press
Utility
>
Memory Catalog
>
More (1 of 2)
>
Security
>
Erase and Overwrite All
>
Confirm Overwrite
.
SRAM
All addressable locations are overwritten with random characters.
CPU Flash
All addressable locations are overwritten with random characters and then the flash blocks are erased.
This accomplishes the same purpose of a chip erase, however, only the areas that are no longer in use are
erased and the factory calibration files are left intact. System files are restored after erase.
DRAM
All addressable locations are overwritten with random characters.
Hard Disk
All addressable locations are overwritten with a single character. (This is insufficient for top secret data,
according to DoD standards. For top secret data, the hard drive must be removed and destroyed.)
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...