652
Chapter 18
Troubleshooting
Basic Signal Generator Operations
Data Storage
Registers With Previously Stored Instrument States are Empty
The save/recall registers are backed-up by a battery when line power to the signal generator is not
connected. The battery may need to be replaced.
To verify that the battery has failed, follow these steps:
1. Turn off line power to the signal generator.
2. Unplug the signal generator from line power.
3. Plug in the signal generator.
4. Turn on the signal generator.
5. Observe the display for error messages.
If either error message
−
311 or
−
700 is stored in the error message queue, the signal generator’s battery
has failed.
6. Refer to the
E4428C/38C ESG Signal Generators Service Guide
for battery replacement instructions.
Saved an Instrument State in a Register but the Register is Empty or Contains the Wrong State
If you have selected a register number that is greater than 99, the signal generator will automatically select
register 99 to save your instrument state.
If the register number you intended to use is empty or contains the wrong instrument state, press the
following keys:
Recall
>
99
>
Enter
.
This recalls register 99. The lost instrument state may be saved there.
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...