Chapter 4
165
Basic Digital Operation
Using Waveform Scaling
5. Press
ARB Setup
>
More (1 of 2)
>
Waveform Runtime Scaling
and adjust the front panel knob or use the
number keys to enter a new value. The new scaling value is instantly applied to the playing waveform.
Runtime scaling adjustments are not cumulative, as the values are always relative to original amplitude
of the waveform file.
In Multitone
1. Press
Preset
>
Mode
>
Multitone
>
Multitone Off On
to On. This generates a multitone arbitrary waveform
for use in this procedure.
2. Press
More (1 of 2)
>
ARB Setup
>
More (1 of 2)
>
Waveform Runtime Scaling
and adjust the front panel knob
or use the number keys to enter a new value. The new scaling value is instantly applied to the playing
waveform. Runtime scaling adjustments are not cumulative, as the values are always relative to original
amplitude of the waveform file.
Scaling a Waveform File in Volatile Memory
This procedure enables you to permanently scale a waveform file. You can then store the scaled waveform
segment to non-volatile memory for future use. Scaling is cumulative and non-reversible.
1. Press
Preset
>
Mode
>
Custom
>
Arb Waveform Generator
>
Digital Modulation Off On
to On. This generates a
custom arbitrary waveform for use in this procedure. You can also use a previously stored or
downloaded waveform.
2. Press
Mode
>
Dual ARB
>
Select Waveform
and ensure that
AUTOGEN_WAVEFORM
is highlighted on the
display.
AUTOGEN_WAVEFORM
is the default name assigned to the waveform you generated in the
previous step.
3. Press
Select Waveform
. This selects the waveform and returns you to the previous softkey menu.
4. Press
ARB Setup
>
Waveform Utilities
and ensure that
AUTOGEN_WAVEFORM
is highlighted on the display.
5. Press
Scale Waveform Data
>
Scaling
>
70
>
%
>
Apply to Waveform
. The waveform is now reduced to 70
percent of its original amplitude. Once this type of scaling is applied to the waveform it cannot be
undone. Repeated scaling applications have a cumulative effect on the waveform.
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...