Chapter 18
649
Troubleshooting
Basic Signal Generator Operations
There are three power search modes: manual, automatic, and span.
When
Power Search
is set to Manual, pressing
Do Power Search
executes the power search calibration routine
for the current RF frequency and amplitude. In this mode, if there is a change in RF frequency or amplitude,
you will need to press
Do Power Search
again.
When
Power Search
is set to Auto, the calibration routine is executed whenever the frequency or amplitude of
the RF output is changed.
When
Power Search
is set to Span, pressing
Do Power Search
executes the power search calibration routine
over a selected range of frequencies at one time. The power search corrections are then stored and used
whenever the signal generator is tuned within the selected range of frequencies.
RF Output Power too Low
1. Look for an
OFFS
or
REF
indicator in the
AMPLITUDE
area of the display.
OFFS
tells you that an amplitude offset has been set. An amplitude offset changes the value shown in the
AMPLITUDE
area of the display but does not affect the output power. The amplitude displayed is equal to
the current power output by the signal generator hardware plus the value for the offset.
To eliminate the offset, press the following keys:
Amplitude
>
More (1 of 2)
>
Ampl Offset
>
0
>
dB
.
REF
tells you that the amplitude reference mode is activated. When this mode is on, the displayed
amplitude value is
not
the output power level. It is the current power output by the signal generator
hardware minus the reference value set by the
Ampl Ref Set
softkey.
To exit the reference mode, follow these steps:
a. Press
Amplitude
>
More (1 of 2)
.
b. Press
Ampl Ref Off On
until Off is highlighted.
You can then reset the output power to the desired level.
2. If you are using the signal generator with an external mixer, see
“Signal Loss While Working with
3. If you are using the signal generator with a spectrum analyzer, see
“Signal Loss While Working with
Spectrum Analyzers” on page 647
.
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...