Chapter 3
53
Basic Operation
Configuring the RF Output
4. Press
RF On/Off.
The display annunciator has changed from
RF OFF
to
RF ON
. The power at the RF OUTPUT connector
is
−
20 dBm.
5. Press
Incr Set
>
10
>
dB
.
This changes the amplitude increment value to 10 dB.
6. Use the up arrow key to increase the output power by 10 dB.
The
AMPLITUDE
area displays
10.00 dB
, which is the power output by the hardware (-20 dBm plus
10 dBm) minus the reference power (
−
20 dBm). The power at the RF OUTPUT connector changes to
−
10 dBm.
7. Press
Ampl Offset
>
10
>
dB
.
This enters a 10 dB offset. The
AMPLITUDE
area displays
20.00 dB
, which is the power output by the
hardware (
−
10 dBm) minus the reference power (
−
20 dBm) plus the offset (10 dB). The
OFFS
indicator
is activated. The power at the RF OUTPUT connector is still
−
10 dBm.
Configuring a Swept RF Output
The signal generator has two sweep types: step and list.
NOTE
List sweep data cannot be saved within an instrument state, but
can
be saved to the memory
catalog. For instructions on saving list sweep data, see
.
During swept RF output, the
FREQUENCY
and
AMPLITUDE
areas of the signal generator’s
display are deactivated, depending on what is being swept.
This section provides an explanation of the differences between step sweep and list sweep. You will learn
two ways to configure the signal generator’s RF output to sweep a defined set of frequency and amplitude
points. You will create a step sweep and then you will use these points as the basis for a new list sweep.
Step Sweep
When a step sweep is activated, the signal generator sweeps the RF output based on the values entered for
RF output start and stop frequencies and amplitudes, a number of equally spaced points (steps) to dwell
upon, and the amount of dwell time at each point. The frequency, amplitude, or frequency
and
amplitude of
the RF output will sweep from the start amplitude/frequency to the stop amplitude/frequency, dwelling at
equally spaced intervals defined by the
# Points
softkey value.
Step sweep provides a linear progression through the start-to-stop frequency and/or amplitude values. You
can toggle the direction of the sweep, up or down. When the
Sweep Direction Down Up
softkey is set to Up,
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...