Chapter 6: Replacing Assemblies
To remove and replace the acquisition board assembly
120
7
Carefully lift the back of the board to separate the acquisition board connectors from
the power distribution board.
8
Lift the board back from the front panel until the BNC connectors clear the front panel,
then lift the board away from the chassis.
Figure 6-20
Separate acquisition board from power distribution board
9
To re-install the acquisition board assembly, reverse the above procedure.
10
Now the calibration factors must be set. See “To set the calibration factors after
replacing the acquisition board” on page 121.
Connectors
Содержание DSO80204B
Страница 7: ...6...
Страница 17: ...Chapter 1 General Information Oscilloscope Dimensions 16...
Страница 34: ...3 To run the self calibration 35 Calibration...
Страница 35: ...34 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 92: ...Chapter 5 Troubleshooting Motherboard Verification 91 Figure 5 14 Sense cable...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 139 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 140 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 141 Power Supply and PC Motherboard...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 142 Sleeve and Accessory Pouch...
Страница 149: ...Chapter 8 Theory of Operation 148 Figure 8 1 Oscilloscope Block Diagram...
Страница 157: ...156...