Chapter 5: Troubleshooting
Front Panel Display Trouble Isolation
86
Front Panel Display Trouble Isolation
Connect external monitor to VGA port. Turn unit on.
Does
display appear
on ext. monitor
?
No
Yes
Is
front Panel
black?
Check display cable
connection to display card
and LCD
No
Does
front panel
display
work?
Try a golden display
card to verify failure
Does
display work
with golden
card?
Replace LCD.
Replace display
card
Replace
motherboard.
Check inverter board control
cable.
Yes
Check voltage on pin 1 of
J3 on display card. Should
be 12V.
Is
voltage
OK?
Replace LCD.
Yes
Does
front panel
display
work?
Replace
inverter.
No
Front Panel Display Trouble Isolation
Go to 'Primary Trouble Isolation'.
Yes
No
Yes
No
No
Yes
Replace
display card.
Содержание DSO80204B
Страница 7: ...6...
Страница 17: ...Chapter 1 General Information Oscilloscope Dimensions 16...
Страница 34: ...3 To run the self calibration 35 Calibration...
Страница 35: ...34 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 92: ...Chapter 5 Troubleshooting Motherboard Verification 91 Figure 5 14 Sense cable...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 139 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 140 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 141 Power Supply and PC Motherboard...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 142 Sleeve and Accessory Pouch...
Страница 149: ...Chapter 8 Theory of Operation 148 Figure 8 1 Oscilloscope Block Diagram...
Страница 157: ...156...
Страница 159: ...158 Index...