Chapter 5: Troubleshooting
Power Supply Trouble Isolation
78
These trouble isolation instructions help isolate the problem to the assembly level when the
power system is not operating. Because of advanced power system protection features, the
problem may not be with the supply itself, and therefore you will need to work through the
procedure systematically to determine the source of the fault.
1
Check the power supply voltages from the power supply. See Figure 5-6 for the location
of these test points. Table 5-7 on page 78 shows the allowable range of power supply
voltages.
Figure 5-6
Power Supply Voltage Test Locations
Table 5-7
Power Supply Voltage Limits
Supply Voltage Specification
Limits
+12.6 V
±
0.1 V
+12.5 V to +12.7 V
-12.6 V
±
0.1 V
-12.5 V to -12.7 V
-5.3 V
±
0.05 V
-5.25 V to -5.35 V
-5.1 V
±
0.05 V
-5.05 V to -5.15 V
-6.1 V
±
0.05 V
-6.05 V to -6.15 V
+5.1 V
±
0.05 V
+5.05 V to + 5.15 V
+3.35 V
±
0.03 V
+3.32 V to +3.38 V
Содержание DSO80204B
Страница 7: ...6...
Страница 17: ...Chapter 1 General Information Oscilloscope Dimensions 16...
Страница 34: ...3 To run the self calibration 35 Calibration...
Страница 35: ...34 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 92: ...Chapter 5 Troubleshooting Motherboard Verification 91 Figure 5 14 Sense cable...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 139 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 140 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 141 Power Supply and PC Motherboard...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 142 Sleeve and Accessory Pouch...
Страница 149: ...Chapter 8 Theory of Operation 148 Figure 8 1 Oscilloscope Block Diagram...
Страница 157: ...156...
Страница 159: ...158 Index...