Chapter 5: Troubleshooting
Acquisition Trouble Isolation
95
Acquisition Trouble Isolation
If Acquisition board has been removed by a prior procedure, reinstall Acquisition board.
Acquisition Trouble Isolation
Does
vertical test
group pass?
Does
trigger test group
pass?
Yes
Does
TimeBase test
group pass?
Yes
Does
ADC test group
pass?
Yes
Does
acquisition memory
test group
pass?
Yes
Does
acquisition interface
test group
pass?
Does
misc. scope
test group
pass?
Yes
Yes
Replace
acquisition
board.
Replace the interface board, A10. If
that does not work, replace the
display board.
Does
temp sense
pass?
Does
fan status
pass?
Check fans
Replace fans
that are not
turning, if
none are
turning
replace
power board
Does
NVram tests
pass?
Replace
interface
card.
Go to 'Primary
Trouble Isolation'.
No
No
Yes
Yes
Replace
acquisition
board.
No
No
No
No
No
No
No
No
Go to 'Primary
Trouble Isolation'.
Yes
Содержание DSO80204B
Страница 7: ...6...
Страница 17: ...Chapter 1 General Information Oscilloscope Dimensions 16...
Страница 34: ...3 To run the self calibration 35 Calibration...
Страница 35: ...34 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 92: ...Chapter 5 Troubleshooting Motherboard Verification 91 Figure 5 14 Sense cable...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 139 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 140 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 141 Power Supply and PC Motherboard...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 142 Sleeve and Accessory Pouch...
Страница 149: ...Chapter 8 Theory of Operation 148 Figure 8 1 Oscilloscope Block Diagram...
Страница 157: ...156...
Страница 159: ...158 Index...