Chapter 4: Testing Performance
Offset Accuracy Test
46
15
Set the channel 1 offset value to 400.0 mV. This can be done using the front panel control
or:
a
Pull down the Setup menu and select Channel 1 or click the Channel 1 setup icon.
b
Click the Offset control arrows to change the offset value or click on the offset value
and enter 400.0 mV in the dialog box.
c
Enter 400.0 mV in the Enter Offset dialog box.
16
Set the Aux Out voltage (V
Aux Out
) to +400.0 mV as follows:
a
Pull down the Utilities menu and select Calibration.
b
Change the Aux Output function to DC (top left corner).
c
Set the Level to 400.0 mV.
d
Click on Close.
17
Press the Clear Display key on the scope, wait for the #Avgs value (top left corner of
screen) to return to 256 and then record the DMM voltage reading as V
DMM+
and the
scope Vavg reading as V
Scope+
in the Offset Gain Test section of the Performance Test
Record.
Channel 1
setup icon
Содержание DSO80204B
Страница 7: ...6...
Страница 17: ...Chapter 1 General Information Oscilloscope Dimensions 16...
Страница 34: ...3 To run the self calibration 35 Calibration...
Страница 35: ...34 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 92: ...Chapter 5 Troubleshooting Motherboard Verification 91 Figure 5 14 Sense cable...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 139 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 140 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 141 Power Supply and PC Motherboard...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 142 Sleeve and Accessory Pouch...
Страница 149: ...Chapter 8 Theory of Operation 148 Figure 8 1 Oscilloscope Block Diagram...
Страница 157: ...156...
Страница 159: ...158 Index...