Chapter 6: Replacing Assemblies
To remove and replace the CD-ROM drive
128
To remove and replace the CD-ROM drive
Use this procedure to remove and replace the CD-ROM drive. When necessary, refer to other
removal procedures.The graphics in this chapter are representative of the oscilloscope at the
time of this printing. Your unit may look different.
1
Disconnect the power cable from the instrument and remove the top sleeve.
2
Remove the following cables:
• SATA Cable
• CD-ROM Power cable
3
Using a Torx T10 driver, remove the three CD-ROM support bracket screws holding
the CD-ROM drive in place.
When re-assembling, torque the three Torx T10 screws to 5 in-lb.
4
Using a T20 driver remove the rear panel screw holding the CD-ROM drive in place.
When re-assembling, torque the Torx T20 screw to 18 in-lb.
5
Move the CD-ROM drive back towards the front frame, then lift and pull to remove.
Figure 6-26
Removing the CD-ROM assembly
Torx
T20
SATA cable
Power
cable
Torx
T10
Torx
T10
Torx
T10
Содержание DSO80204B
Страница 7: ...6...
Страница 17: ...Chapter 1 General Information Oscilloscope Dimensions 16...
Страница 34: ...3 To run the self calibration 35 Calibration...
Страница 35: ...34 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 92: ...Chapter 5 Troubleshooting Motherboard Verification 91 Figure 5 14 Sense cable...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 139 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 140 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 141 Power Supply and PC Motherboard...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 142 Sleeve and Accessory Pouch...
Страница 149: ...Chapter 8 Theory of Operation 148 Figure 8 1 Oscilloscope Block Diagram...
Страница 157: ...156...
Страница 159: ...158 Index...