Chapter 5: Troubleshooting
To install the fan safety shield
67
Default Setup
A Default Setup is provided to return the oscilloscope to a known state. The default
setup can be used to undo previous setups so that they do not interfere with the current
measurement. Use the default setup when a procedure requires it.
• Press the Default Setup key to set the oscilloscope to the default state.
To install the fan safety shield
1
Disconnect the oscilloscope power cord and remove the cover.
If necessary, refer to the procedures in chapter 6 "Replacing Assemblies".
2
Clip the fan safety shield over the outside of the oscilloscope chassis next to the fans.
Figure 5-1
Installing the Fan Safety Shield