Agilent B1542A User’s Guide, Edition 4
5- 21
PLSDIV Test Definitions
PLSDIV IdVg
PLSDIV IdVg
This test definition controls SMU (source monitor unit) to force the drain voltage,
controls PGU (pulse generator) to force the gate voltage pulse, controls the
oscilloscope to monitor the gate voltage and the drain voltage, calculates the drain
current, and displays the Id-Vg characteristics.
Figure 5-4
Pulsed IV Id-Vg Measurement Result Example
Device Parameters
The following parameters are available to set the test condition.
Polarity
Polarity of source output. Nch (force the specified value) or Pch
(force the negative specified value). See “Parameters” on page
5-31.
Lg
Gate length, in m.
≥
1 nm, 4 digits or 1 nm resolution.
Wg
Gate width, in m.
≥
1 nm, 4 digits or 1 nm resolution.
Temp
Temperature, in degree. 4 digits or 0.001 degree resolution.
IdMax
Maximum drain current measurement range, in A. 0 to 80 mA,
4 digits or 1
μ
A resolution.
DeviceDelay
Delay time caused by device under test, in s.
-
1
μ
s to 1
μ
s,
4 digits or 100 ps resolution. See “Parameters” on page 5-31.
,G
9G
9J
9G6WRS
9G6WDUW9J6WHS
9G6WDUW
Содержание B1542A
Страница 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Страница 21: ...2 Installation ...
Страница 49: ...3 Performing System Setup and Compensation ...
Страница 65: ...4 Performing Measurement ...
Страница 79: ...5 PLSDIV Test Definitions ...
Страница 115: ...6 PLSDIV TIS Commands ...
Страница 139: ...7 Status Code and Error Messages ...