5- 20
Agilent B1542A User’s Guide, Edition 4
PLSDIV Test Definitions
PLSDIV IdVd
Test Output
The following output parameters are displayed after the test.
Id
Drain current, in A.
Vd
Target drain voltage, in V.
VdSet
Drain voltage, in V. SMU output value.
VdInt
Drain voltage, in V. Oscilloscope monitor value.
VdConv
Convergence result of the drain voltage automatic adjustment.
Vg
Target gate voltage, in V.
VgMon
Gate voltage, in V. Oscilloscope monitor value.
VgConv
Convergence result of the gate voltage automatic adjustment.
IdPerWg
Drain current per unit gate width, in A/
μ
m.
IdPerWg = Id / Wg
VgMon, VdInt, and Id values will be the value obtained at the data extraction timing
on the pulse top. See Figure 5-9.
VdConv and VgConv returns 0 (convergence condition is satisfied) or
-
1 (not
satisfied). If the automatic adjustment is OFF, 0 is returned.
If the drain voltage automatic adjustment function is ON, VdInt will be close to Vd.
And if the function is OFF, VdSet will be close to Vd.
If the gate voltage automatic adjustment function is ON, VgMon will be close to Vg.
And if the function is OFF, Vpgu-out will be close to Vg.
Figure 5-3
Test Output Parameters
'
6
*
6KXQWUHVLVWRU=
9G6HW
9G,QW 9G6HW=,G
,G
9J0RQ
9SJXRXW
3*8
608
6XE
Содержание B1542A
Страница 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Страница 21: ...2 Installation ...
Страница 49: ...3 Performing System Setup and Compensation ...
Страница 65: ...4 Performing Measurement ...
Страница 79: ...5 PLSDIV Test Definitions ...
Страница 115: ...6 PLSDIV TIS Commands ...
Страница 139: ...7 Status Code and Error Messages ...