3- 6
Agilent B1542A User’s Guide, Edition 4
Performing System Setup and Compensation
Starting System Setup
Figure 3-2
PLSDIV Setup Test Definition
14. Set the DcSwitch field as follows.
•
Yes for the DC + pulsed IV test system or the option B1542A-024
•
No for the pulsed IV test system
15. Save the test setup to your preset group (My Favorite Setup).
16. Connect GPIB cable between the system controller and the system instruments
if it is still not connected.
17. Connect BNC cable between the pulse generator TRIGGER OUT connector and
the oscilloscope AUX Trig In connector if it is still not connected.
18. Turn on the system instruments.