2- 26
Agilent B1542A User’s Guide, Edition 4
Installation
Rack-mounting Pulse/dc Switch Units
Rack-mounting Pulse/dc Switch Units
Rack-mount pulse/dc switch units as shown below. The required parts are included
in the option B1542A-025. See Figure 2-20.
Required parts:
Procedure:
1. Assemble the rail kits
B
and
C
as shown in Figure 2-19. Then make sure and use
the screw hole available for your system rack.
2. Secure the rail assemblies
B
and
C
to the proper position of the system rack by
using the typical rack-mounting method of the support rail kit.
3. Assemble the pulse/dc switch unit panel by mounting the following units on the
panel
A
. Use screws listed above to fix the units.
4. Align the pulse/dc switch unit panel (
A
with
D
,
E
, and
F
) with the rail
assemblies secured to the system rack in the step 2, gently slide the panel on the
rails until it stops, and secure the panel to the system rack by using the typical
rack-mounting method of the rack-mount flange kit.
Reference
Designator
Agilent Part No.
Quantity
Description
A
B1542-00201
1
Panel
B
B1542-01211
1
Rail kit
C
B1542-01212
1
Rail kit
-
0515-0372
12
Screw
Reference
Designator
Agilent Part No.
Description
D
B1542-60001
Switch control distributor
E
B1542-60002
Drain pulse/dc switch unit
F
B1542-60003
Gate pulse/dc switch unit
Содержание B1542A
Страница 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Страница 21: ...2 Installation ...
Страница 49: ...3 Performing System Setup and Compensation ...
Страница 65: ...4 Performing Measurement ...
Страница 79: ...5 PLSDIV Test Definitions ...
Страница 115: ...6 PLSDIV TIS Commands ...
Страница 139: ...7 Status Code and Error Messages ...