Agilent B1542A User’s Guide, Edition 4
5- 19
PLSDIV Test Definitions
PLSDIV IdVd
Extended Test
Parameters
The following parameters are available to set the test condition. For more details,
see “Parameters” on page 5-31.
TransTime
Pulse leading/trailing edge transition time, in s. 2 ns to 500 ns, 3
digits or 1 ns resolution.
MeasTime
Measurement timing at the pulse top. 0 to 1, 3 digits or 0.01
resolution.
SmplNum
Oscilloscope sampling number. 16 to 262144, 6 digits or 1 step.
SmthNum
Oscilloscope smoothing number. 3 to 4001, 4 digits or 1 step.
SmthStat
Status of the oscilloscope smoothing. Enable or Disable.
AvgNum
Oscilloscope averaging number. 1 to 4095, 4 digits or 1 step.
VdAdjust
Status of the drain voltage automatic adjustment. Enable or
Disable.
VgAdjust
Status of the gate voltage automatic adjustment. Enable or
Disable.
RangingMode
Measurement ranging mode. Auto or Fixed.
ConnectChk
Status of the connection check at initialization. Enable or
Disable.
GateZero
Status of the gate monitor channel zero measurement at
initialization. Enable or Disable.
DcSwitch
Availability of the pulse/dc switch units. Yes (exists, pulsed IV
measurement connection will be made) or No (none).
SWAddress
GPIB address of the 11713A switch controller. Integer, 1 to 32.
Содержание B1542A
Страница 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Страница 21: ...2 Installation ...
Страница 49: ...3 Performing System Setup and Compensation ...
Страница 65: ...4 Performing Measurement ...
Страница 79: ...5 PLSDIV Test Definitions ...
Страница 115: ...6 PLSDIV TIS Commands ...
Страница 139: ...7 Status Code and Error Messages ...