![Agilent Technologies B1542A Скачать руководство пользователя страница 74](http://html.mh-extra.com/html/agilent-technologies/b1542a/b1542a_user-manual_2868004074.webp)
4- 10
Agilent B1542A User’s Guide, Edition 4
Performing Measurement
Pulsed Id-Vd Measurement
Pulsed Id-Vd Measurement
Pulsed Id-Vd measurement can be performed as follows.
Figure 4-8
PLSDIV IdVd Test Definition
1. Open the PLSDIV IdVd test definition on the EasyEXPERT application test
environment.
2. Set the measurement condition to the Device Parameters, Test Parameters, and
Extended Test Parameters. For the description of the entry fields, see “PLSDIV
IdVd” on page 5-17.
3. Connect the device under test (MOSFET) as shown in “RF Probes” on page 2-4.
4. Click the Single button at the upper right corner of the EasyEXPERT screen.
Pulsed Id-Vd measurement is started.
Figure 4-9 shows the pulsed Id-Vd measurement result example.
Содержание B1542A
Страница 1: ...Agilent Technologies Agilent B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 20: ...1 12 Agilent B1542A User s Guide Edition 4 Introduction Accessories and Options ...
Страница 21: ...2 Installation ...
Страница 49: ...3 Performing System Setup and Compensation ...
Страница 65: ...4 Performing Measurement ...
Страница 79: ...5 PLSDIV Test Definitions ...
Страница 115: ...6 PLSDIV TIS Commands ...
Страница 139: ...7 Status Code and Error Messages ...