-151
Invalid string data
An invalid character string was received. Check to see if you have
enclosed the character string in single or double quotes and that the
string contains valid
ASCII
characters.
-158
String data not allowed
A character string was received but is not allowed for the command.
Check the list of parameters to verify that you have used a valid
parameter type. Example:
BMOD:NCYC ’TEN’
-161
Invalid block data
Applies only to the
DATA:DAC VOLATILE
command. For a definite-
length block, the number of bytes of data sent does not match the
number of bytes that you specified in the block header. For an
indefinite-length block, an
EOI
(End-or-Identify) was received without
an accompanying <
new line
> character.
-168
Block data not allowed
Block data was received but is not allowed for the command. Check to
see if you have sent the correct data type with the command.
-170 to -178
Expression errors
The function generator does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (
GET
) or
*TRG
was received but the trigger
was ignored. Make sure that you have selected the correct trigger source.
5
Chapter 5 Error Messages
Execution Errors
231
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents Contents 12 ...
Страница 15: ...1 Quick Start 1 ...
Страница 31: ...2 Front Panel Menu Operation 2 ...
Страница 55: ...3 Features and Functions 3 ...
Страница 126: ...124 ...
Страница 127: ...4 Remote Interface Reference 4 ...
Страница 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Страница 228: ...226 ...
Страница 229: ...5 Error Messages 5 ...
Страница 244: ...242 ...
Страница 245: ...6 Application Programs 6 ...
Страница 272: ...270 ...
Страница 273: ...7 Tutorial 7 ...
Страница 299: ...8 Specifications 8 ...
Страница 304: ...302 ...