![Agilent Technologies 33120A Скачать руководство пользователя страница 193](http://html.mh-extra.com/html/agilent-technologies/33120a/33120a_user-manual_2867471193.webp)
*SAV {0|1|2|3}
Store up to four different instrument configurations.
[
Stored in non-volatile memory
]
Four memory locations (numbered 0, 1, 2, and 3)
are available to store
instrument configurations. The state storage feature “remembers” the
function (including arbitrary waveforms), frequency, amplitude,
dc offset, duty cycle, as well as any modulation parameters. To recall
a stored state, you must use the same memory location used
previously to store the state.
From the remote interface only, you can use memory location “0” to
store a fourth instrument configuration (you cannot store to this
memory location from the front panel). However, keep in mind that
memory location “0” is automatically overwritten when the power is
turned off.
Any arbitrary waveforms downloaded to “
VOLATILE
” memory
are not
remembered. However, if an arbitrary waveform is being output
from
non-volatile
memory when the state is stored, the waveform
data is stored. The stored waveform is output when the instrument
state is recalled.
If you delete an arbitrary waveform after storing the state, the
waveform data is lost and the function generator will not output the
waveform when the state is recalled. The “
SINC
” waveform is output
in place of the deleted waveform.
When power is turned off, the function generator automatically stores
its state in memory location “0”. You can configure the function
generator to automatically recall the power-down state when power is
restored. The recall mode is
disabled
when the function generator is
shipped from the factory.
Select the
POWER ON LAST STATE
command from the
SYS MENU
to
enable the power-down recall mode. Select
POWER ON DEFAULT
to
disable the power-down recall mode.
See “Power-Down Recall Mode”
on page 109 for more information.
4
Chapter 4 Remote Interface Reference
System-Related Commands
191
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents Contents 12 ...
Страница 15: ...1 Quick Start 1 ...
Страница 31: ...2 Front Panel Menu Operation 2 ...
Страница 55: ...3 Features and Functions 3 ...
Страница 126: ...124 ...
Страница 127: ...4 Remote Interface Reference 4 ...
Страница 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Страница 228: ...226 ...
Страница 229: ...5 Error Messages 5 ...
Страница 244: ...242 ...
Страница 245: ...6 Application Programs 6 ...
Страница 272: ...270 ...
Страница 273: ...7 Tutorial 7 ...
Страница 299: ...8 Specifications 8 ...
Страница 304: ...302 ...