To Determine When a Command Sequence is Completed
1
Send a device clear message to clear the function generator’s
output buffer (e.g.,
CLEAR 710
).
2
Clear the event registers with the
*CLS
(clear status) command.
3
Enable the “operation complete” bit (bit 0) in the Standard Event
register by executing the
*ESE 1
command.
4
Send the
*OPC?
(operation complete query) command and enter the
result to ensure synchronization.
5
Execute your command string to program the desired configuration,
and then execute the
*OPC
(operation complete) command as the last
command. When the command sequence is completed, the “operation
complete” bit (bit 0) is set in the Standard Event register.
6
Use a serial poll to check to see when bit 5 (standard event) is set in the
Status Byte summary register. You could also configure the function
generator for an
SRQ
interrupt by sending
*SRE 32
(Status Byte enable
register, bit 5).
Chapter 4 Remote Interface Reference
The SCPI Status Registers
206
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents Contents 12 ...
Страница 15: ...1 Quick Start 1 ...
Страница 31: ...2 Front Panel Menu Operation 2 ...
Страница 55: ...3 Features and Functions 3 ...
Страница 126: ...124 ...
Страница 127: ...4 Remote Interface Reference 4 ...
Страница 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Страница 228: ...226 ...
Страница 229: ...5 Error Messages 5 ...
Страница 244: ...242 ...
Страница 245: ...6 Application Programs 6 ...
Страница 272: ...270 ...
Страница 273: ...7 Tutorial 7 ...
Страница 299: ...8 Specifications 8 ...
Страница 304: ...302 ...