You can set the units for output amplitude to Vpp, Vrms, or dBm.
See “Output Units” on page 64 for more information.
For
dc volts
, the output level is actually controlled by setting the
offset voltage. You can set the dc voltage to any value between
5 Vdc
into 50 ohms or
10 Vdc into an open circuit.
See “DC Offset Voltage”
on page 62 for more information.
To select
dc volts
from the front panel, press
Offset
and
hold it down
for more than 2 seconds
.
Front-Panel Operation
: To set the output amplitude, press
Ampl
.
Then use the knob, arrow keys, or “Enter Number” mode to set the
amplitude.
Remote Interface Operation
:
VOLTage {<
amplitude
>|MINimum|MAXimum}
You can also use the
APPLy
command to select the function,
frequency, amplitude, and offset with a single command.
3
Chapter 3 Features and Functions
Output Configuration
61
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents Contents 12 ...
Страница 15: ...1 Quick Start 1 ...
Страница 31: ...2 Front Panel Menu Operation 2 ...
Страница 55: ...3 Features and Functions 3 ...
Страница 126: ...124 ...
Страница 127: ...4 Remote Interface Reference 4 ...
Страница 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Страница 228: ...226 ...
Страница 229: ...5 Error Messages 5 ...
Страница 244: ...242 ...
Страница 245: ...6 Application Programs 6 ...
Страница 272: ...270 ...
Страница 273: ...7 Tutorial 7 ...
Страница 299: ...8 Specifications 8 ...
Страница 304: ...302 ...