Using the APPLy Command
See also “Output Configuration,” starting on page 55 in chapter 3.
The
APPLy
command provides the most straightforward method to
program the function generator over the remote interface. You can
select the function, frequency, amplitude, and offset all in one command.
For example, the following statement outputs a 3 Vpp sine wave at
5 kHz with a -2.5 volt offset using
APPLy
:
"APPL:SIN 5 KHZ, 3.0 VPP, -2.5 V"
The syntax statements for the
APPLy
command are shown on page 143.
Output Frequency
For the
frequency
parameter of the
APPLy
command, the output
frequency range depends on the function currently selected.
You can substitute “
MIN
imum”, “
MAX
imum”, or “
DEF
ault” in place of
a specific value for the
frequency
parameter.
MIN
selects the lowest
frequency allowed for the selected function.
MAX
selects the highest
frequency allowed.
The default frequency for all functions is 1 kHz
.
Function
Sine
Square
Ramp
Triangle
Built-In Arbs 1
Minimum Frequency
100
m
Hz
100
m
Hz
100
m
Hz
100
m
Hz
100
m
Hz
Maximum Frequency
15
MHz
15
MHz
100
kHz
100
kHz
5
MHz
1 There are five built-in arbitrary waveforms stored in non-volatile memory:
sinc, negative ramp, exponential rise, exponential fall, and cardiac.
Chapter 4 Remote Interface Reference
Using the APPLy Command
138
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents Contents 12 ...
Страница 15: ...1 Quick Start 1 ...
Страница 31: ...2 Front Panel Menu Operation 2 ...
Страница 55: ...3 Features and Functions 3 ...
Страница 126: ...124 ...
Страница 127: ...4 Remote Interface Reference 4 ...
Страница 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Страница 228: ...226 ...
Страница 229: ...5 Error Messages 5 ...
Страница 244: ...242 ...
Страница 245: ...6 Application Programs 6 ...
Страница 272: ...270 ...
Страница 273: ...7 Tutorial 7 ...
Страница 299: ...8 Specifications 8 ...
Страница 304: ...302 ...