As shown below, the 33120A has a fixed output source resistance of
50 ohms. During calibration, output amplitudes are calibrated for both
the open-circuit voltage (no load) and the terminated output voltage
(loaded). The terminated output amplitude is calibrated for an exact
50 ohm load. Since the function generator’s output resistance and the
load resistance form a voltage divider, the measured output voltage of
the function generator will vary with load resistance value and accuracy.
When the function generator’s output is loaded with a 0.2% accuracy
termination, an additional (negligible) 0.2% amplitude error is created.
Using a 5% accuracy termination will add 5% additional error to
specified output amplitudes.
If the function generator’s output is measured with no load connected,
the output will be approximately
twice
the displayed amplitude
(V
gen
instead of V
load
)
. In some applications, you might continually use
the function generator in a “no-load” conditions. In such applications,
remembering to
double
the function generator’s displayed amplitude can
cause many errors. The 33120A allows you to specify the function
generator’s load condition using the
OUTPUT:LOAD
command; thus
enabling the function generator to display the correct output amplitude.
50
W
50
W
V
load
Vgen
7
Chapter 7 Tutorial
Output Amplitude Control
281
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents Contents 12 ...
Страница 15: ...1 Quick Start 1 ...
Страница 31: ...2 Front Panel Menu Operation 2 ...
Страница 55: ...3 Features and Functions 3 ...
Страница 126: ...124 ...
Страница 127: ...4 Remote Interface Reference 4 ...
Страница 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Страница 228: ...226 ...
Страница 229: ...5 Error Messages 5 ...
Страница 244: ...242 ...
Страница 245: ...6 Application Programs 6 ...
Страница 272: ...270 ...
Страница 273: ...7 Tutorial 7 ...
Страница 299: ...8 Specifications 8 ...
Страница 304: ...302 ...