The Standard Event register is cleared when:
You execute the
*CLS
(clear status) command.
You query the event register using the
*ESR?
(event status register)
command.
The Standard Event enable register is cleared when:
You execute the
*ESE 0
command.
You turn on the power and have previously configured the function
generator using the
*PSC 1
command.
The enable register
will not
be cleared at power-on if you have
previously configured the function generator using
*PSC 0.