To Secure Against Calibration
You can secure the function
generator against accidental calibration either from the front panel or
over the remote interface.
The function generator is secured when
shipped from the factory and the security code is set to “HP033120”.
Be sure to read the security code rules on page 118 before attempting to
secure the function generator.
Front-Panel Operation
:
1: UNSECURED
(CAL MENU)
If the function generator is unsecured, you will see the above
command when you go into the
CAL MENU
. To secure the function
generator, go to the “
PARAMETER
” level of the
UNSECURED
command, enter the security code, and then press
Enter
.
When you go to the “
COMMANDS
” level in the
CAL MENU
again,
you will notice that the function generator is secured. Notice also that
the “
CALIBRATE
” command is now hidden and you cannot perform a
calibration.
Remote Interface Operation
:
CALibration:SECure:STATe {OFF|ON},<
code
>
To secure the function generator, send the above command with the
same code as used to unsecure calibration. For example,
"CAL:SEC:STAT ON,HP033120"
¾
000000:CODE
1:
SECURED
Chapter 3 Features and Functions
Calibration Overview
120
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents Contents 12 ...
Страница 15: ...1 Quick Start 1 ...
Страница 31: ...2 Front Panel Menu Operation 2 ...
Страница 55: ...3 Features and Functions 3 ...
Страница 126: ...124 ...
Страница 127: ...4 Remote Interface Reference 4 ...
Страница 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Страница 228: ...226 ...
Страница 229: ...5 Error Messages 5 ...
Страница 244: ...242 ...
Страница 245: ...6 Application Programs 6 ...
Страница 272: ...270 ...
Страница 273: ...7 Tutorial 7 ...
Страница 299: ...8 Specifications 8 ...
Страница 304: ...302 ...